Chapter 4

Supervisor Menu

86

zoll.com

9650-001358-01 Rev. B

P3 Mean Alarm

The following options allow you to set the limits of the pediatric P3 Mean alarm.

Upper Limit Enable

The Upper Limit Enable parameter allows you to enable or disable the upper limit.

The values for the Upper Limit Enable parameter are:

Lower Limit Enable

The Lower Limit Enable parameter allows you to enable or disable the lower limit.

The values for the Lower Limit Enable parameter are:

Upper Limit

The Upper Limit parameter allows you to set the upper limit.

The values for the Upper Limit parameter are:

Lower Limit

The Lower Limit parameter allows you to set the lower limit.

The values for the Lower Limit parameter are:

Default Neonate

The following options allow you to specify the alarm limits for neonate patients.

HR/PR Alarm

The following options allow you to set the limits of the neonate HR/PR alarm.

Upper Limit Enable

The Upper Limit Enable parameter allows you to enable or disable the upper limit.

The values for the Upper Limit Enable parameter are:

Possible Values:

On, Off

Default:

Off

Possible Values:

On, Off

Default:

Off

Possible Values:

-28-300

Default:

110

Possible Values:

-30-298

Default:

50

Possible Values:

On, Off

Default:

Off