Parameter Menus

The following table includes parameter ranges and default values for the Parameter menus:

Menu

Submenu

Parameter Name

Default Value

Parameter Values

HR/PR

HR/PR Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

Adult: 120

Pediatric: 150

Neonate: 200

32-300

Lower Limit

Adult: 50

Pediatric: 50

Neonate: 100

30-300

Extreme Brady/Tachy
Alarms

Tachy: +20

Brady: -10

Tachy: +5-+30 by 5’s

Brady: -5--30 by 5’s

LTA Monitoring

Off

On, Off

HR/PR Tone

On

On, Off

HR/PR Selected
Source

ECG

ECG, SpO2, P1, P2, P3

ECG Sweep Speed

25 mm/s

25 mm/s, 50 mm/s

NIBP

NIBP Systolic Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

Adult: 220

Pediatric: 145

Neonate: 100

Adult: 32-260

Pediatric: 32-160

Neonate: 22-120

Lower Limit

Adult: 75

Pediatric: 75

Neonate: 50

Adult: 30-258

Pediatric: 30-158

Neonate: 20-118

NIBP Diastolic Alarm Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

Adult: 110

Pediatric: 100

Neonate:70

Adult: 22-220

Pediatric: 22-130

Neonate: 12-110

Lower Limit

Adult: 35

Pediatric: 35

Neonate: 30

Adult: 20-218

Pediatric: 20-128

Neonate: 10-108

NIBP MAP Alarm

Upper Limit Enable

Off

On, Off

Lower Limit Enable

Off

On, Off

Upper Limit

Adult: 120

Pediatric: 110

Neonate: 80

Adult: 25-230

Pediatric: 25-140

Neonate: 15-110