Nibp parameter settings, Nibp systolic alarm – ZOLL X Series Monitor Defibrillator Rev B User Manual

Page 31

Advertising
background image

NIBP Parameter Settings

9650-001358-01 Rev. B

ZOLL X Series Configuration Manual

25

NIBP Parameter Settings

The NIBP parameter settings allow you to view and modify functions related to using the X
Series unit for NIBP monitoring, and includes:

NIBP Systolic Alarm

NIBP Diastolic Alarm

NIBP MAP Alarm

NIBP Mode

NIBP Auto Mode Interval

SmartCuf On/Off

NIBP Systolic Alarm

The following options allow you to set the limits of the NIBP Systolic Alarm for all patient age
groups.

Upper Limit Enable

The Upper Limit Enable parameter allows you to enable or disable the upper limit.

The values for the Upper Limit Enable parameter are:

Lower Limit Enable

The Lower Limit Enable parameter allows you to enable or disable the lower limit.

The values for the Lower Limit Enable parameter are:

Upper Limit

The Upper Limit parameter allows you to set the upper limit.

The values for the Upper Limit parameter are:

Possible Values:

On, Off

Default:

Off

Possible Values:

On, Off

Default:

Off

Possible Values:

Adult: 32-260

Pediatric: 32-160

Neonate: 22-120

Default:

Adult: 220

Pediatric: 145

Neonate: 100