ZOLL X Series Monitor Defibrillator Rev B User Manual

Page 81

Advertising
background image

Alarms Parameters

9650-001358-01 Rev. B

ZOLL X Series Configuration Manual

75

Lower Limit

The Lower Limit parameter allows you to set the lower limit

.

The values for the Lower Limit parameter are:

No Breath Alarm

The No Breath Alarm parameter allows you to specify the time of the pediatric No Breath
alarm.

The values for the No Breath Alarm parameter are:.

SpO2 Alarm

The following options allow you to set the limits of the pediatric SpO2 alarm/

Upper Limit Enable

The Upper Limit Enable parameter allows you to enable or disable the upper limit.

The values for the Upper Limit Enable parameter are:

Lower Limit Enable

The Lower Limit Enable parameter allows you to enable or disable the lower limit.

The values for the Lower Limit Enable parameter are:

Upper Limit

The Upper Limit parameter allows you to set the upper limit.

The values for the Upper Limit parameter are:

Lower Limit

The Lower Limit parameter allows you to set the lower limit.

The values for the Lower Limit parameter are:

Possible Values:

0-145

Default:

3

Possible Values:

Off, 10 sec, 15 sec, 20 sec, 25 sec, 30 sec, 40 sec, 50 sec, 60 sec

Default:

Off

Possible Values:

On, Off

Default:

Off

Possible Values:

On, Off

Default:

Off

Possible Values:

52-100

Default:

100

Possible Values:

50-98

Default:

85