Hardware Checks

If all LEDs are still OFF 10 or more seconds after plugging the 82357A USB
cable into a USB port, and all Windows Plug and Play Manager activity and
Connection Expert refreshes have ceased, start your troubleshooting
sequence by performing hardware checks. If any action taken results in a
change in the LED status, go to “Software Installation Checks” or to
“Software Configuration Checks”.

Check USB Cables, USB Interface, Host PC

Begin the hardware checks by checking connections between the 82357A
and PC (plus USB hubs, if used).

1

Check USB Cable Connections.

Check the 82357A USB cable for

a good connection to the USB port on the PC or on the USB hub.
If you are using a USB hub, verify that the hub is connected to the PC.

2

Unplug/replug the 82357 USB cable.

If this does not change the

LED status, try plugging the 82357A into another USB port.

3

Check PC USB Port.

Verify that the PC USB port is functional and

powered (such as checking by using another USB device).

4

Check PC State.

Verify that host computer is not in a Suspended

power management state.

5

Check USB Hub

. Try disconnecting the 82357A from the hub and

connecting it directly to a USB port on the PC. Some USB hubs are
vulnerable to static shock.

6

Check USB Cables for Damage.

Check the USB cable for cuts/crushes.

Since the end connectors are somewhat fragile, check for bent/
misaligned/crushed connectors.

Reboot the PC

If Steps 1, 2, 3, 4, 5, or 6 do not change the LED status, reboot the PC.