Chapter 3

53

Troubleshooting the 82357A

Software Configuration Checks

1

Unplug/Replug the 82357A.

If unplugging and replugging the 82357A

causes the

Agilent 82357A USB/GPIB Interface Detected

dialog box to

appear or the Connection Expert window to refresh, the problem is
solved. If not, go to Step 2.

2

Shut Down and Restart IO Control.

Take these steps to shut down and

then restart the IO Control. Taking these actions should result in all
attached and configured 82357As to be initialized and to display only
the green Ready LED

a If the blue IO icon is displayed, click the icon and then click

Exit

.

A dialog box explaining the consequences of removing the IO
Control appears. Click

OK

to shut down the IO Control.

b If the blue IO icon is not displayed, either the icon display has been

turned off or the IO Control (and associated iproc82357.exe and
iprocsvr.exe) are not active. In this case, select

Start | Programs |

Agilent IO Libraries Suite

and click

IO Control

to re-start the

IO Control and display the blue IO icon.

Check USB Scanner

In general, USB scanners do not cause problems with the 82357A.
However, if you do have problems with 82357A operation and have a
scanner installed on your system that uses a USB port, unplug the scanner
and then plug the 82357A into the port.

If the 82357A will configure without your scanner attached to your system,
the scanner was locking out the 82357A from using the USB bus. In this
case, contact your scanner’s manufacturer to request software or firmware
updates for the scanner.

NOTE

Rebooting your PC should ALWAYS restart the IO Control and
iprocsvr.exe and iproc82357.exe

.