2.6 Power Supply Voltage and

Load Resistance

When configuring the loop, make sure that the

external load resistance is within the range in the

figure below.

(Note) With an intrinsically safe transmitter, external load

resistance includes safety barrier resistance.

600

250

0

10.5 16.6

25.2

42

External

load

resistance

R (Ω)

Power supply voltage E (V DC)

F0203.ai

Communication

applicable range

R= E–10.5

0.0244

Figure 2.3

Relationship between Power Supply

Voltage and External Load Resistance