1 checking contact output -13, Caution – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 101

Advertising
background image

IM 11M12A01-03E

7-9

7. Startup

Parameter: Humidity
Set range

Min. humidity conc:

0 %H2O

Max. humidity conc:

25 %H2O

Output damping: 0 s

Output mode: Linear

Enter

mA-output1 range

Oxygen

Humidity
Mixing

F7.10.1.EPS

Parameter: Mixing
Set range

Min. mixing ratio:

0. 000 kg/kg

Max. mixing ratio:

0. 200 kg/kg

Output damping: 0 s

Output mode: Linear

Enter

mA-output1 range

F7.10.2.EPS

Figure 7.10.1 mA-output1 Range Figure 7.10.2 Mixing Ratio Display
Selection

7.7.2

Minimum Current (4 mA) and Maximum Current (20 mA) Settings

To set the analog output range, follow these steps:

(1) Select the Setup from the Execution/Setup display.
(2) From the Commissioning (Setup) display, select the “mA-output setup.” Select “mA-

output1” from the “mA-outputs” display. The “mA-output1 range” that displays
“Parameter: Humidity,” as shown in Figure 7.10 then appears.

(3) To set the minimum humidity at 4 mA, choose the Min. humidity conc. To set the

maximum humidity at 20 mA, choose the Max. humidity conc.

(4) To set 50% H

2

O, type in 050 and press the [Enter] key.

(5) Set “mA-output2” in the same manner as in the appropriate steps above.

CAUTION

Note: Analog output range settings should be limited. For more details, consult Section
8.1, “Current Output setting,” later in this manual.