2 output hold setting, 1 definition of equipment status, Default values -8 – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 118

Advertising
background image

IM 11M12A01-03E

8-6

8.2

Output Hold Setting

The “output hold” functions hold an analog output signal at a preset value during the
equipment’s warm-up time or calibration or if an error arises. Outputs 1 and 2 can be set
individually. Table 8.2 shows the analog outputs that can be retained and the individual
states.

Table 8.2

Equipment status

During warm-up Under

Under calibration On Error

maintenance During blow back occurrence

Output hold
values available

4 mA

20 mA

Without hold

feature

Retains output

from just before

occurrence

Set value

(2.4 to 21.6 mA)

:

The output hold functions are available.

T8.2E.EPS

8.2.1

Definition of Equipment Status

(1) During Warm up

“Warming up” is the time required after applying power until the sensor tempera-
ture stabilizes at 750

ЊC, and the equipment is in the measurement mode.

(2) Under maintenance

“During maintenance” is the time, starting from pressing the Setup key in the basic
panel display the Execution/Setup display, to go back to the basic panel
display. The display panels under the Execution/Setup display, which are in Figure

7.3.1, are included.

(3) Under calibration” (see Chapter 9, “Calibration”)

“During calibration” is the time, during manual calibration, starting from entering
the Manual calibration display (shown in Figure 8.1), and making a series of
calibrations, until the Calibration End key is pressed or until the preset output
stabilization time elapses.

Enter

Open span gas valve.

Set flow span gas to

600ml/min.

Valve opened

Cancel calibration

Manual calibration

F8.1E.EPS

Figure 8.1