5 input contact settings, 1 input contact functions, Default values -17 – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 128

Advertising
background image

IM 11M12A01-03E

8-16

8.5

Input Contact Settings

8.5.1

Input Contact Functions

The converter input contacts execute set functions by accepting a remote contact signal.
Table 8.8 shows the functions executed by a remote contact signal.

Table 8.8 Input Contact Functions

Item

Function

Calibration-gas press. low

While the contact signal is on, neither semi-automatic nor

automatic calibration is possible.

Measuring range change

While contact input is On, range of Analog Output 1 is switched as

follows: When analog output 1 range is set to "Humidity", then output

range is switched to 0 to 100% H

2

O. When analog output 1 range is

set to "Mixing ratio", then output range is switched to 0 to 1 kg/kg.

When analog output 1 range is set to "Oxygen", then range is switched

to 0 to 25% O

2

. While range is switched by the contact input,

[Range] is displayed on the screen. See Figure 8.12.

Calibration start

If the contact signal is applied, semi-automatic calibration

starts (only if the semi-automatic or automatic mode has been

setup). Calibration is started with an applied one- to 11-second

time interval single-output contact signal. Even if a continuous

contact signal is applied, a calibration is not repeated.

If you want to perform a second calibration,

turn the contact signal off and then back on.

Process

If the contact signal is on, heater power will be switched off.

upset

(A one- to 11-second time interval single-output signal is

available as a contact signal.) If this operation starts, the sensor

temperature decreases and an error occurs. To restore it to normal,

turn the power off and then back on, or reset the analyzer.

Blow back start

If the contact signal is on, blow-back starts. (A one- to 11-second

time interval single-output signal is available as a contact signal.)

Even if a continuous contact signal is applied, calibration is

not repeated. If you want to make a second calibration,turn the contact

signal off and then back on. (Refer to Section 10.2, "Blow back.")

T8.8E.EPS

Tag:

0.0

%H

2

O

4.00mA -Output1
4.00mA -Output2

Range

F8.12E.EPS

Figure 8.12 Changing Measuring Range with Input Contact