Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 163

Advertising
background image

IM 11M12A01-03E

10-15

10. Other Functions

Item

Data to be initialized

Default setting

Parameter

Oxygen

Alarm setting

concentration

Hysteresis

0.1% O

2

Delayed action of alarm contact

3 seconds

High-high alarm

None

Alarm value

100% O

2

Alarm set value High-limit alarm

None

Alarm value

100% O

2

Low-limit alarm

None

Alarm value

0% O

2

Low-low alarm

None

Alarm value

0% O

2

Humidity alarm Hysteresis

0.1% H

2

O

setting

Contact delay 3

seconds

Alarm data

High high alarm

Off

Humidity alarm Alarm set value

100% H

2

O

set value

High alarm

Off

Alarm set value

100% H

2

O

Low alarm

Off

Alarm set value

0% H

2

O

Low low alarm

Off

Alarm set value

0% H

2

O

Mixing-ratio

Hysteresis

0.01 kg/kg

alarm setting

Contact delay

3 seconds

Mixing-ratio

High high alarm

Off

alarm set value Alarm set value

1 kg/kg

High alarm

Off

Alarm set value

1 kg/kg

Low low alarm

Off

Alarm set value

0 kg/kg

Low low alarm

Off

Alarm set value

0 kg/kg

T10.5E-2.EPS

to be continued next page