Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 154

Advertising
background image

IM 11M12A01-03E

10-6

10.1.2.2

Sampling Period

To plot a graph, set the sampling period for the measurement data.

This graph allows the plotting of 60 data items on one graduation on the time axis. So, if
you set a ten-second sampling period, one graduation corresponds to 600 seconds
(Figure 10.5). The allowable sampling periods range from 1 to 30 seconds. If you set a
one-second sampling period, the axis of the abscissas then corresponds to five minutes.
If you set it to 30 seconds, the axis of the abscissas then corresponds to 150 minutes.

0.0%O

2

25.0%O

2

10min./div

12.3%O

2

60 data items

Upper limit set

Lower limit set

Time per graduation,
calculated by the set
sampling period

Currently measured value

Sampling period

F10.5E.EPS

Figure 10.5 Plotting Graph for Sampling Period

10.1.2.3 Setting Upper and Lower Limit Values on Graph

Set upper- and lower-limit values on the graph in the following procedure:

Press Upper limit in the Trend graph display. The numeric-data entry key appears. Enter
the upper-limit value. Also enter the lower-limit value in the same way. The allowable
settings for both upper-limit and lower-limit values range from: (1) 0 to 100 %O

2

for

oxygen concentration; (2) 0 to 100 %H

2

O for humidity; and (3) 0 to 1 kg/kg for mixing

ratio.

Advertising