Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 21

Advertising
background image

IM 11M12A01-03E

2-2

Repeatability: (Note1)

Ϯ1 vol% H

2

O (sample gas pressure 2 kPa or less)

Linearity:

(Excluding standard gas tolerance)

(Note1)

(Use oxygen of known concentration (in the measuring range) as the
zero and span calibration gas.)
Ϯ2 vol% H

2

O; (Sample gas pressure: within

Ϯ0.49 kPa)

Ϯ3 vol% H

2

O; (Sample gas pressure: 2 kPa or less)

Drift:

(Excluding the first two weeks in use)

(Note1)

both zero and span

Ϯ3 vol% H

2

O/month

Response Time: Response of 90 % within 5 seconds. (Measured after gas is introduced

from calibration-gas inlet and analog output starts changing.)

(Note1) These tolerances do not apply to the pressure compensated
version, or where natural convection is used for the reference air.