Figure b max. and min. humidity set ranges – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 114

Advertising
background image

IM 11M12A01-03E

8-2

Humidity (amount-of-moisture-content) setting range

The minimum humidity is set to 0% H

2

O or ranges from 26 to 100% H

2

O. The maxi-

mum humidity ranges from 25% to 100% H

2

O, and must be greater than 0.8 times plus

23 the humidity set for the minimum.

Setting example 1

If the setting (for a 4 mA current) is 0% H

2

O, you must set the maximum (20 mA)

point at more than 25% H

2

O.

Setting example 2

If the setting (for a 4 mA current) is 26% H

2

O, you must set the maximum (20 mA)

point at more than 44% H

2

O, (26

ϫ 0.8 + 23% H

2

O). (Numbers after the decimal

point are rounded up.)

FBE.EPS

Minimum humidity (amount-of-moisture-content) (for a 4-mA current), % H

2

O

Maximum humidity (amount-of-moisture-content)

(for a 20-mA

current), % H

2

O

40

50

60

70

80

90

100

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

0

0

0

Range over which humidity
can be set

Outside ranges

Figure B Max. and Min. Humidity Set Ranges