6 pressure at sampling gas outlet, 7 example configuration of gas sampling system, Sensor (11) no – Yokogawa IR400 User Manual

Page 21: Co, co, And o

Advertising
background image

3 - 8

IM 11G02N01-01E

(1) Gas extractor

15

Drain

Exhaust

2-way
solenoid
valve

Air

ZERO NO

SO

2

CO

O

2

CO

2

NO
/N

2

SO

2

/N

2

CO

2

/N

2

CO
/N

2

O

2

/N

2

Infrared gas analyzer

main unit CO,CO

2

(O

2

)

IR400

or larger

(10)
Zirconia
O

2

senor

φ

10/

φ

8

Teflon
tube

(ZX8D)

(2) Mist filter

(3) Safety
drain

trap

(4) Gas aspirator

Electronic
gas cooler

(5)

(8) Flowmeter

(7) Membrane filter

(7) Membrane filter

(11)
NO

2

/NO

converter

(6)
Solenoid
valve

Pressure
reducing valve

(9) Standard gas

Infrared gas analyzer

main unit NO, SO

2

Description

Gas extractor with a heating
type stainless steel filter of stan-
dard mesh 40 m

Removes drain, mist, and dust.

The safety drain trap divided
into two rooms for positive and
negative pressure. It monitors
and adjusts the sample gas pres-
sure.

For aspiration of sample gas

Dries the moisture in sample
gas to a dew point of approx.
2˚C.

Used for introducing calibra-
tion gas.

PTFE filter used to eliminate
fine dust particles and permit
monitoring of dust adhering
condition on the front panel of
the gas analyzer.

Name

(8) Flowmeter

(9) Standard gas

(10) Zirconia O

2

sensor

(11) NO

2

/NO con-

verter

Description

Adjusts and monitors the flow
rate of sample gas.

Reference gas used for calibrat-
ing zero and span of the ana-
lyzer. Total 6 cylinders required
for zero gas air, span gas NO,
SO

2

, CO, CO

2

and O

2

.

External zirconia oxygen
sensor used for measuring the
oxygen concentration in
sample gas.
(This is not necessary in case
when O

2

sensor is built-in.)

Added to NOx analyzer.
A special catalyst material for
efficient conversion of NO

2

gas to NO is used.

Name

(1) Gas extractor

(2) Mist filter

(3) Safety drain trap

(4) Gas aspirator

(5) Electronic gas

cooler

(6) Solenoid valve

(7) Membrane filter

3.4.6 Pressure at sampling gas outlet

Pressure at the sampling gas outlet should be adjusted to atmospheric pressure.

3.4.7 Example configuration of gas sampling system

The following illustrates a typical system configuration for five component gas measurement for
monitoring combustion exhaust gas from boiler, refuse incinerator, etc.

Contact Yokogawa for system configuration matching the particular use or further information.