IM 11G02N01-01E

6.2 Calibration setting

This mode is used to set calibration concentration and actions. The calibration setting involves
calibration concentration, zero calibration, calibration range and auto calibration component.

6.2.1 Setting of calibration concentration

It allows you to set concentrations of the standard gas (zero and span) of each channel used for
calibration.

MODE

(

)

ENT

(

)

ENT

(

)

ENT

(1) During measurement, press the

MODE

key to

display the User mode.

(2) Point the cursor to “Calibration

Parameters” by pressing the

or

key.

Press the

ENT

key.

(3) In the “Calibration Parameters” screen

that appears, point the cursor to “Calibra-

tion Value” by pressing the

or

key. Press the

ENT

key.

(4) In the “Calibration Concentration Ch

Selection” screen that appears, point the
cursor to Ch you want to set by using the

or

key. Press the

ENT

key.