Solenoid valve drive signal output for calibration – Yokogawa IR400 User Manual

Page 29

Advertising
background image

3 - 16

IM 11G02N01-01E

Connector <CN3>

Connector <CN3> provides outputs in combination with calibration action during auto calibration
and manual calibration.

An output is from a transistor (ratings: 5 V/50 mA).

A transistor is turned ON before starting each calibration.

Sample selection output is ON during measurement and OFF during calibration.

If calibration is not performed, the other transistors are OFF.

In case of auto calibration, sequential output is ON/OFF according to the setting.

Refer to “Item 6.4 Setting of auto calibration”.

Note) No. 9 pin is for solenoid valve ON/OFF relay drive power (5 V DC/0.5 A, max).

Use No. 9 with reference to the diagram.

Connector

<CN3>

(Transister output)

<D-sub 9P>

Contact output for

zero calibration

Contact output for

sample gas selection

Contact output for Ch1

span calibration

Contact output for Ch2

span calibration

Contact output for Ch3
span calibration
Contact output for Ch4
span calibration
Contact output for Ch5
span calibration

Power for relay drive
5V DC

1

2

3

4

5

6

7

8

9

Solenoid valve drive signal output for calibration

Cable

I/O terminal module

Relay board

Contact relay

CN3

5V

SV1 to SV4: solenoid valves

Solenoid valve
drive power

SV1

SV2

SV3

SV4

1

3

4

5

9

Standard gas for calibration

<Gas sampling system>

Refer to Item 3.4.7 Example configuration
of gas sampling system .

<Electrical system>

Example of using solenoid valve drive signal output for calibration

Relay board and exclusive cable (D-sub 9p straight cable: 1.5 meters)
are available on request.

SV4

Ch2

span

SV3

Ch1

span

SV2

SV1

Zero

Membrane filter

Electronic
gas cooler

Flowmeter