(1) Turn off the power switch, stop the

sample gas, and allow the zero gas to
flow for several minutes to purge the
cell interior.

Loosen the setscrew (2 pieces) from
the top cover and remove it.

(2) Remove the internal gas inlet tube.

(3) Loosen the 2 detector set bolts.

Note) The distribution cell, block cell

and detector are fastened by the
same bolts.

(4) Using the furnished cell mounting tool,

turn the window fixture to the left and
remove it from the cell. (See the
structure of sample cell (block cell) in
Fig. 7-2.)

(5) For cleaning the infrared transmission

window and cell inside surface, first
eliminate coarse dust by soft brush or
the like and then wipe them by soft rag.

The window is easy to get scratched.
Pay utmost attention so as not to
damage it.

(6) After the end of sample cell cleaning,

mount the cell in place and proceed to
running.

After cleaning sample cell, be sure to
perform optical zero adjustment (see
Item 7.3.3) and moisture interference
compensation adjustment (see Item
7.3.4).

Caution

Infrared transmission window

Inner O ring

Outer O ring

O ring holder

Window fixture

Top cover

F10.12E.eps

Gas inlet tube set cap nut

Detector set bolt

If the window or the cell interior is very dirty, use a soft cloth moistened with absolute alcohol.
A slightly corroded infrared transmission window or sample cell can be remedied by gently rubbing
with chromium oxide powder on cleaning cloth but an excessively corroded one must be replaced.
When cleaning, do not exert an excessive stress.

7.3.2 Cleaning method for sample cell (block cell)