6 - 8

IM 11G02N01-01E

(5) On the “calibration range selection”

screen that appears, select “both” or
“current” by pressing the

or the

key.

• If “both” is selected, zero or span

calibration is performed with Range 1
and Range 2 of the selected Ch
interlocked.

• If “current” is selected, zero or span

calibration is performed only for the
range displayed when calibration of
selected Ch is performed.

Press the

ENT

key after the selection, and

the specified calibration is performed.

To perform calibration for “both,” set the same calibration gas concentration for both

ranges.

Note

When setting NOx and CO to “both”

Two cursors will appear in both ranges (Ch1 and Ch4).

Manual Calibration screen

(

)

ENT

End of Calibtation Range Setting

Ch1

NOx

Ch2

SO

2

Range 1: 0 to 100 ppm

Range 2: 0 to 2000 ppm

Range 1: 0 to 100 ppm
Range 2: 0 to 2000 ppm

both

current

Ch1: Range 1 and Range 2 are calibrated together with zero and span calibration.

Ch2: Only currently displayed range is calibrated with zero and span calibration.

Example

To close “Setting of Calibration Range” or

to cancel this mode midway, press the

ESC key. A previous screen will return.

To close “Setting of Calibration Range”

ESC