5 setting of auto zero calibration, 1 autozero calibration, Description of setting items – Yokogawa IR400 User Manual

Page 59

Advertising
background image

6 - 20

IM 11G02N01-01E

6.5 Setting of auto zero calibration

6.5.1 Autozero calibration

Auto zero calibration is automatically carried
out at the time when zero calibration is set.
Components for which a calibration is to be
made are determined by setting of auto cali-
bration component in Item 6.2.4.

Before changing the setting of auto zero
calibration, set the ON/OFF to OFF.

(1) During measurement, press the

MODE

key

to display the User mode.

(2) Point the cursor to “Setting of Auto Zero

Calibration” by pressing the

or

key. Press the

ENT

key.

(3) In the “Setting of Auto Zero Calibration”

screen that appears, point the cursor to
any item you want to set by pressing the

or

key. Press the

ENT

key.

(4) In the “Auto Zero Calibration Parameter

Setting” screen that appears, perform the
value entry or the setting. For the value
entry or setting change, use the

or

key. To change the setting, use the
key to move the cursor to the right.

After setting, press the

ENT

key, and auto

zero calibration is carried out by the
entered setting value.

(

)

ENT

MODE

(

)

ENT

ENT

End of Auto Zero Calibration Setting

Description of setting items

• Start Time : Setting at the first calibration

(day of the week, hour, minute)

• Cycle

: A period between the start time of one

calibration and another
(unit : hour/day)

• Flow Time : The time required for the calibration gas

to be replaced in the cell

• ON/OFF

: ON/OFF of auto zero calibration

To close the “Setting of Auto Zero Calibration”
or cancel this mode midway, press the key.
A previous screen will return.

To close "setting of Auto Zero Calibration"

ESC