Yokogawa IR400 User Manual

Page 87

Advertising
background image

7- 6

IM 11G02N01-01E

7.3.3 Optical zero adjustment method (optical balance adjustment)

The adjustment is performed at reassembly after removing the sample cell, etc. for cleaning, etc.

If the following operation is maladjusted, the measurement may adversely be affected. If you are
not trained for adjustment, do not carry out this operation but contact the distributor or our service-
man.

Caution

No. of components

to be measured

1-component meter

3-component meter

4-component meter

2-component meter

a

Main

NO-SO

2

CO

2

-CO

NO-CO

CO

2

Main

NO

Main

NO-SO

2

-CO

* O

2

is excluded from the number of components.

* Main is signal input value from the main detector of each component.

Comp is signal input value from interference compensation detector of each component.

If low range exceeds the range of 0 to 10vol%, detector signal of comp is not usable.

NO-SO

2

-CO

2

-CO

b

Comp

CO

2

Comp

NO

Comp

c

CO

Main

SO

2

Main

d

CO

Comp

SO

2

Comp

e

f

g

h

NO

Main

NO

Comp

CO

Main

CO

Comp

NO

Main

NO

Comp

CO

Main

CO

Comp

SO

2

Main

SO

2

Comp

NO

Main

NO

Comp

CO

2

Main

CO

Main

CO

Comp

SO

2

Main

SO

2

Comp

Sensor values of which are not included in measuring components should be ignored.

<Correspondence between measurement detector and indicated position>

(1) Remove the top cover. Allow dry N

2

or air to flow through the analyzer unit
sample gas inlet until the reading
stabilizes. The sample gas is introduced
directly to the INLET of analyzer unit
through the gas cylinder.

(2) Proceed to an optical adjustment in the

maintenance mode. The display on the
operation panel of the main unit is as
illustrated on the right. Balance adjust-
ment is not required if the display falls
within

±

100.

a

b

c

d

e

f

g

h

Piping at optical balance adjustment

Short

Open

Introduce directly from cylinder

(not required
for measuring
unit 1 only)

Optical system 2
(measuring unit 2)

Optical system 1

(measuring unit 1)

Advertising