Environmental conditions, Mounting and shape – Yokogawa PR300 Power and Energy Meter User Manual

Page 65

Advertising
background image

A-8

IM 77C01E01-01E

Appendix 1 Specifications of PR300

Environmental Conditions

Normal operating conditions

Warm-up time

At least 30 minutes

Ambient temperature

0 to 50

°C (reference temperature: 23Ϯ2°C)

Temperature change

10

°C/h or less

Ambient humidity

20 to 90% RH (no condensation)

Magnetic field

400A/m or less

Continuous vibration

10 to 60 Hz, 0.035 mm, 75 minutes; 60 to 150 Hz, 4.9 m/s

2

, 75 minutes

Short-time vibration

14.7 m/s

2

for 15 seconds or less

Shock

98 m/s

2

or less (for shock time of 11 ms)

Installation

Indoor installation only

Mounting position

Vertical surface mounting only

Installation altitude

2000 m or less

Effect on operating conditions

Effects of ambient temperature

Analog output:

Ϯ0.05% of F.S./°C or less

Effects on power supply voltage variations

Analog output:

Ϯ0.05% of F.S./°C or less

Transport and storage conditions

Temperature

-20 to 70

°C

Humidity

5 to 95% RH (no condensation)

Shock and dropping of package

90 cm (provided that an external packing box is used)

Mounting and Shape

Materials

Casing

Polycarbonate resin (PC), UL94 V-0

Terminal block

Polybutylene terephthalate (PBT), UL94 V-0

Terminal cover

Polyamide resin (PA6), UL94 V-2

Mounting method

Panel mounting

Connection Method

M3 screws for terminal connections (analog output, pulse output, demand alarm output, digital
input, and RS-485 communication)

M4 screws for terminal connections (voltage/current input and power supply)

RJ45 connection (Ethernet communication, when the Ethernet communication function is specified)

External Dimensions (including a terminal cover)

110

ϫ 110 ϫ 128 mm (H ϫ W ϫ D)

96

ϫ 96 ϫ 126 mm (H ϫ W ϫ D)

Weight

Approx. 600 g (when accessories such as mounting bracket are attached)