Data format UPP® (Universal Pyrometer Protocol) 29

7 Data format UPP® (Universal Pyrometer

Protocol)

Via interface and suitable communication software or via “Test” function of the InfraWin
(see 3.3.5 Basic settings

Test) commands can be exchanged directly with the pyrometer.

The data exchange occurs in ASCII format with the following transmission parameters:

The data format is: 8 data bits, 1 stop bit, even parity (8,1,e).

The device responds to the entry of a command with: output (e.g. the measuring value) + CR
(Carriage Return, ASCII 13), to pure entry commands with “ok” + CR.
Every command starts with the 2-digit device address AA (e.g. “00”). This is followed by 2 small

command letters (e.g. “em” for level of emissivity

e

), finished with CR.

This is followed, if necessary for that command, by the ASCII parameter “X”. If this parameter “X” is
omitted, then the device resets with the current parameter.

A „?“ after the small command letters answers with the respective settings (only at setting
commands, not at inquiry commands).

Example: Entry: “00em“ + CR

The emissivity setting (

e

) of the device with the address 00 is returned

Answer: “0970“ + CR means Emissivity = 0.97 or 97.0%

Description

Command

Parameters

Reading temperature
value:

AAms

Output: XXXXX (dec., in

1

/

10

°C or °F)

last digit is the decimal place (88880 = Temp.-Overflow)

Reading temperature
value repeated:

AAmsXXX

XXX = 000…999 (XXX = number of measuring values)

Reading basic
temperature range:

AAmb

Output: XXXXYYYY (hex 8-digit, °C or °F)

XXXX

=

beginning

of

temp.

range

YYYY=

end

of

temp.

range

Reading temperature sub
range:

AAme

same as mb

Reading emissivity:

AAem

Output: XXXX (dez. 0010 … 1000 in ‰)

Emissivity:

AAemXX

XX = (10…99%), 00 = 100% (decimal)

Emissivity:

AAemXXXX

XXXX = (0010 … 1000‰) (decimal)

Exposition time t

90

:

AAezX

X = 0 … 6 (decimal)

0 = intrinsic time constant

1 = 0.01 s

3 = 0.25 s

5 = 3.00 s

2 = 0.05 s

4 = 1.00 s

6 = 10.00 s

External deletion:

AAlx

Simulation of an external deletion contact

Delete maximum value:

AAlzX

*

)

X = 0 … 8 (dez.)
0 = Maximum value storage off
1

=

0.01

s

4

=

1.00

s

7

=

external

deletion

2 = 0.05 s

5 =

5.00 s

8 = automatically deletion

3 = 0.25 s

6 = 25.00 s

Setting of temperature
sub range:

AAm1XXXXYYYY

XXXX (hex 4-digit) beginning of temp. range (°C)
YYYY (hex 4-digit) end of temp. range (°C)

Analog output:

AAasX *

)

X = 0 or 1;

0 = 0 to 20 mA

1 = 4 to 20 mA