Configuring Video Settings 19

Example: On the left, the hot object is outside of the selected region, so it is not used in

adjusting the brightness and contrast. On the right, the object is now in the selected region and
the brightness and contrast adjust for a better display.

2.7 Auto-Cal

Automatic calibration (Auto-Cal) is used to keep the camera image
quality looking good. If automatic calibration is active, the camera will
calibrate when the temperature of the camera changes or the timer is

exceeded. The timer can be adjusted in the number box.

Note: Auto-Cal is not available for change without entering the password using
Advanced from the Help pull down menu. See section 4.1 for more information about

entering the password.

Note: The default time is 300 seconds.

To turn on automatic calibration:

1. Check the box under Auto-Cal.
2. Set the timer for calibration (15-7200)

3. Click Save.

The image will freeze while the camera is calibrating. This freeze is sometimes undesirable, so
automatic calibration may be turned off.

To turn off automatic calibration:

1. Uncheck the box under Auto-Cal.

2. A warning prompt will appear. Click Yes to continue disabling the camera or No to

revert back to enabling auto-cal.

3. Click Save.