Caution – HT instruments HT7052 User Manual

Page 29

Advertising
background image

HT7052

EN - 27

5. Press START/STOP key to activate the insulation measurement
6. Wait for the end of the set Timers. The result of test is shown at display (see Fig. 39 or

Fig. 41) with meaning of items descript in Table 8

7. Wait for the object under test to discharge
8. For saving the result see § 7

Parameter at display

Description

Fil0 (Fil1, Fil2, Fil3)

Filter type enabled (see § 6.3)

5000V

Set test voltage – step 125 V

U=5302V

Applied test voltage

I=133nA

Applied test current

39.7G

Result of insulation measurement

C=0.0nF Capacitance

of measured object

Tm:05min 00s

Actual test duration

R1000V=40.0G

Last result of 1

st

step

R2000V=40.0G

Last result of 2

nd

step

R3000V=40.0G

Last result of 3

rd

step

R4000V=39.8G

Last result of 4

th

step

R5000V=39.7G

Last result of 5

th

step

U1=1076V 1

st

step voltage

U2=2141V 2

nd

step voltage

U3=3238V 3

rd

step voltage

U4=4278V 4

th

step voltage

U5=5302V 5

th

step voltage

Table 8: Meaning of parameters of insulation measurement with step voltage

CAUTION

 Timer information is displayed from the start of the measurement until the

completion of each step measurement

 A high-voltage warning symbol appears on the display during the

measurement to warn the operator of a potentially dangerous test voltage

 The value of the capacitance is measured during the final discharge of the

test object

Advertising