10 TECHNICAL SPECIFICATIONS

Accuracy is given as [%rdg + (number of dgt) * resolution] at reference indicated in § 0

INSULATION RESISTANCE MEASUREMENT

Measurement range

Resolution

Accuracy

120k

  999k 1k

(5.0rdg + 3dgt)

1.00M

  9.99M 0.01M

10.0M

  99.9M 0.1M

100M

  999M 1M

1.00G

  9.99G

0.01G

10.0G

  99.9G 0.1G

100G

  999G 1G

1.00T

  10.00T 0.01T

(15.0 rdg + 3 dgt)

FS value of insulation resistance is defined as:

RFS = 1G

 * Utest [V]

Nominal test voltage:

500

 10kV DC

Current capability of test generator:

> 1mA

Short-circuit test current:

5mA

 10%

Automatic discharge of tested object:

yes

Measurement range test voltage

Resolution

Accuracy

0

 9999V

1V

(3.0 rdg + 3V)

 10kV

0.1kV

3.0 rdg

Nominal test voltage:

500

 10kV DC programmable in step of 25V

Accuracy of test voltage:

-0 / +10% + 20V

Output power:

10W max

Measurement range test current

Resolution

Accuracy

0.00

 9.99nA

0.01nA

(5.0 rdg + 0.05nA)

10.0

 99.9nA

0.1nA

100

 999nA

1nA

1.00

 9.99A 0.01A

10.0

 9.99A 0.1A

100

 999A 1A

1.00

 5.50mA

0.01mA


Noise current rejection (resistive load)

Filter option

Maximum current @ 50Hz (mA rms)

Fil0 1.5
Fil1

2.5

Fil2

4.5

Fil3

5

Diagram Test voltage – Resistance

0

2

4

6

8

10

12

0,1

1

10

100

[ M ]

[ k

V

]

Utest=10kV

Utest=5kV