Caution – HT instruments HT7052 User Manual

Page 30

Advertising
background image

HT7052

EN - 28

6.8 WITHSTANDING VOLTAGE TEST
6.8.1 Setting of parameters
1. Switch on the instrument by pressing the ON/OFF key
2. Select with arrow keys  or  the item “WITHSTANDING VOLTAGE DC” on main

menu and confirm with SELECT key. The screen of Fig. 43 is shown by the meter

Fig. 43: Initial screen withstanding test

Fig. 44: Final screen withstanding test

3. Press again the SELECT key to enter in the setup parameters section. The screen of

Fig. 45 is shown by the meter

Fig. 45: Setting parameters

4. Use the arrow keys  or  for the selection of parameters. The herewith Table 9

shows the meaning of the measurement parameters

5. Set the values by using the arrow keys  or . Press SELECT key to select possible

sub-parameters and repeat the settings

6. Press ESC key to save the settings and back to the measurement screen or the

START/STOP key to exit from the settings menu and activate the test

Parameter Description

Ustart

Start test voltage – Range 500V

10kV step 25V

Ustop

Stop test voltage – Range 500V

10kV step 25V

Itrigg

Set trigger leakage current – Range 0.001mA

5mA step 10A

Tstep

Duration of test voltage per one step

Tend

Duration of constant test voltage after reaching stop value

Table 9: Setting of internal parameters

CAUTION

 Tstep and Tend are independent timers. The maximum time for each timer

is 30 min 60 s. Tend begins after the completion of the ramp period. Ramp
period can be calculated from:

Ttot-ramp

 Tstep * [(Ustop-Ustart) / 25V]

 If Tstep is set to 00min 00s, then the ramp voltage increases by

approximately 25 V every 2s