HT instruments HT7052 User Manual

Page 31

Advertising
background image

HT7052

EN - 29

6.8.2 Perform the measurement
1. Switch on the instrument by pressing the ON/OFF key
2. Select with arrow keys  or  the item “WITHSTANDING VOLTAGE DC” on main

menu and confirm with SELECT key. The screen of Fig. 43 is shown by the meter

3. Connect the red part of the Test lead 1 or Test leads 2 (see § 4.2) to the +OUTPUT

input and the black part of the Test leads 2 (see § 4.2) to the –OUTPUT input

4. Connect the tip of Test lead 1 or Test leads 2 (positive) and the black cable of Test

leads 2 (negative) to the object on test (see Fig. 46)

Fig. 46: Connection of instrument for withstanding test

5. Press START/STOP key to activate the measurement
6. Wait until the set timers run out or until breakdown occurs. The result of test is shown

at display (see Fig. 44) with meaning of items descript in Table 10

7. Wait for the object under test to discharge
8. For saving the result see § 7

Parameter at display

Description

2000V

Start test voltage

7000V

Stop test voltage

7221V

Applied test voltage

I=0.002mA

Measured leakage current

Tm:01min 00s

Timer information

Table 10: Meaning of parameters of withstanding test

CAUTION

 Breakdown is detected when the measured current reaches or exceeds the

set current level Itrigg (see § 6.8.1)

 The timer shows the time needed to complete each step during the

measurement and it shows the total measurement period after the
completion of the measurement

 A high-voltage warning symbol appears on the display during the

measurement to warn the operator of a potentially dangerous test voltage

Advertising