Calibrate the ORP Sensor

1. Tap the Calibration icon

to access a list of sensors that are available for

calibration.

2. Tap ORP Sensor.

3. Tap 1-Point Calibration

4. Make sure the vented cap is installed on the calibration cup. Fill the cup to the fill

line with calibration standard. Place the instrument into the calibration cup, and tap
Start.