14 care and maintenance, Maintenance schedule, User-serviceable parts – In-Situ smarTROLL Multiparameter Handheld Operators Manual User Manual

Page 67: O-rings, Rdo fast sensor cap replacement, Ph/orp sensor replacement, Instrument storage, Chapter 11—care and maintenance, Care and maintenance

Advertising
background image

67

Care and Maintenance

Maintenance Schedule

For best results, send the instrument to the manufacturer for factory calibration every 12
to 18 months.

User-Serviceable Parts

The user-serviceable parts on the instrument include the O-rings, the pH/ORP sensor,
and the RDO Sensor Cap.

O-rings

The instrument has several O-rings that can be maintained by the user in order to keep
moisture from entering the instrument and damaging the electronics. Apply a very thin
layer of vacuum grease to new O-rings upon installation. The O-rings are located in the
following areas.

1

Connector

2

Instrument housing

3

pH sensor

4

RDO Sensor

RDO Fast Sensor Cap Replacement

The RDO Fast Sensor Cap has a 1-year typical life (15 months of total usage) after the
sensor takes its first reading, or 36 months from the date of manufacture. Follow the
instructions included in the RDO Sensor Cap Replacement Kit. Replacement caps are
available from In-Situ Inc. or your authorized In-Situ distributor.

pH/ORP Sensor Replacement

To replace the pH/ORP sensor or to refill the reference junction, follow the instructions
in the pH/ORP Sensor Instruction Sheet that is included with the replacement sensor.

Instrument Storage

To store the probe for a week or less, place the probe in the calibration cup with a few
drops of clean water to maintain a moist storage environment.

Advertising
This manual is related to the following products: