The dissolved oxygen sensor cannot be calibrated using the
Quick-Cal procedure.

4. Tap OK.

5. Fill the calibration cup to the fill line with Quick-Cal solution.

6. Place the instrument into the calibration cup, and tap Start.

7. When the calibration is stable, tap the Accept button.