TROLL 9500 Operator’s Manual

0095110 rev. 007 01/09

For DEPTH:

1 Select Pressure in Parameters list, select Edit

2 Display Mode: Select Depth, click Next.

3 Select Depth/Level Conversion: Specific Gravity 1.0.

Click Finish.

QUICK SUMMARY OF PRESSURE SETUP

For SURFACE WATER MONITORING (install probe before
setting these):

Display Mode

Level - Surface

Depth/Level Conversion

Specific Gravity 1.0

Level Reference

Current elevation or gauge
height of water surface

Reference Reading

Now (with probe installed)

For AQUIFER TESTING:

1 Select Pressure in Parameters list, select Edit

Display Mode
Depth/Level Conversion

Specific Gravity 1.0

2 Display Mode: Select Level - Top of Casing, click Next.

3 Select Depth/Level Conversion: Specific Gravity 1.0. Click Next.

4 Select Level Reference 0 & Reference Reading Start of Test.

Click Finish.

After setting the pressure mode, select the UNITS you want:

7IN 3ITU

Pocket-Situ: Tap Home at top of screen, tap Setup at bottom of screen, scroll down to LEVEL/DEPTH, select unit, click OK

2

3

4

4

SECTION 7: MONITORING PRESSURE (WATER LEVEL)