In-Situ TROLL 9500 Operators Manual User Manual

Page 71

Advertising
background image

64

TROLL 9500 Operator’s Manual

0095110 rev. 007 01/09

SN32072
Files:

File Name

SN32072 2006-04-17 171530 TURB.html

SN32072 2006-06-18 102309 DO.html

SN32072 2006-08-14 134200 QuickCal.html

SN32072 2006-11-25 121530 PH.html

SN32072 2007-01-11 154322 Cond.html

Folders:

Folder

SN32072

SN45025

Example of a calibration report index

Click a folder to see its listings

Click a tile to open it

MP TROLL 9500

serial number

Cal Date

Cal Time

(hhmmss)

Parameter

The calibration reports are accessible from the Tools Menu and the
Show Calibration Report button

on the toolbar. They may also

be accessed like other files through Windows Explorer (desktop PC)
or File Explorer (PDA); they are not displayed in the Data Folder. They
may be viewed or printed to provide a complete calibration history.

TIP: Here’s how to find the calibration report indexes:

Desktop or laptop PC—Calibration Reports subfolder in the folder
where Win-Situ 4 is installed

PDA—Calibration Reports subfolder in the folder where Pocket-Situ
4 is installed

HOW OFTEN TO CALIBRATE

No sensor will remain in calibration forever. The calibration frequency
is almost completely determined by the chemical properties of the
mUID

When a sensor or instrument is new, we recommend checking the
readings often (say, once a day) to get an idea of the stability of the
sensor.

#HANGES

The table below may be used as a very general guideline to how long
sensors may be expected to remain in calibration under optimum
conditions:

pH, ORP

1-2 months

Conductivity 2-3

months

D.O. (polarographic) 2-4 weeks

D.O. (optical, RDO) up to a year if foil is not damaged

ISEs

1

day

TIP: For additional information on calibration schedules, see
the Technical Note on Instrument Calibration in the

Downloads section at www.in-situ.com.

HOW TO CHECK IF A SENSOR IS STILL IN CALIBRATION

Immerse the sensor in a calibration standard of known value and at
the same temperature as the original calibration. Compare the sensor
reading to the solution value. Some drift is to be expected, but gener-
ally the readings should fall within the sensor’s accuracy specification.
If readings fall outside the accuracy specification by an amount that
is not acceptable for your current application, recalibration is recom-
mended. You will quickly learn by experience how often you need to
recalibrate a given sensor based on usage.

TIP: Quick Cal solution may be used for a quick check of
pH, ORP, and conductivity. Refer to the values printed on the
label.

SECTION 10: MONITORING WATER QUALITY: OVERVIEW

When you select the sensor . . .

. . . information
on the last
calibration is
shown

pH calibration information for a combination pH/ORP sensor in port 1

Advertising
This manual is related to the following products: