KYORITSU 2009R User Manual

Page 12

Advertising
background image

̶ 10 ̶

●Operating System

ΔΣ modulation

●Display

Liquid crystal display with a maximum

count of 4200 plus annunciators

●Overrange Indication

"OL" is shown on the display

●Response Time

Approx. 2 seconds.

●Sample Rate

About 3 times per second.

●Location for use

Indoor use, Altitude up to 2000m

●Temperature and Humidity 23 +/-5℃ , relative humidity up to 75%

 for Guranteed Accuracy

without condensation

●Operating Temperature

0-40℃ , relative humidity up to 85%

 and Humidity

without condensation

●Storage Temperature

-20-60℃ , relative humidity up to 85%

 and Humidity

without condensation

●Power Source

Two 1.5VDC R6P (SUM-3) batteries

●Current Consumption

Approx. 40mA max (ACA).

●Auto-power-off function

Automatically powered down in about

10 minutes after the last switch

operation

(power consumption: about 200μA)

●Overload Protection

DC/AC current ranges:

  2400A AC for 10sec

DC/AC voltage ranges:

  1200V AC/DC for 10sec

Resistance range:

  1000V AC/DC for 10sec

●Withstand Voltage

6880V AC, 50/60Hz for 5 seconds

between electrical circuit and housing

case or metal part of the jaws

●Insulation Resistance

10MΩ or greater at 1000V between

electrical circuit and housing case or

metal part of the jaws

●Conductor Size

Approx. 55㎜ diameter max.

●Dimensions

250(L)x105(W)x49(D)mm

●Weight

About 540g

●Accessories

Test leads M-7017

R6P battery

Carrying case M-9094

Instruction manual

Recorder output Plug M-8201

●Optional Accessories

Multi-Tran M-8008

Output Lead M-7014.

Advertising