KYORITSU 2009R User Manual

Page 27

Advertising
background image

̶ 25 ̶

NOTE
 ● If the instrument in the Data Hold mode goes into "Auto-power-

off," it will return to the normal mode.

7-3 LoHz Function

  In ACV or ACA range, if frequency of the voltage or current under

test is 40Hz or lower, the display indicates "LoHz".

 "LoHz" is also indicated where input is 3% of range or less.

7-4 OUTPUT Terminal (For current ranges only)

DANGER

●Never use the instrument on a circuit above 750VAC/1000VDC.
  This may cause electrical shock hazard and damage to the

instrument or the circuit under test.

● Do not make measurement with the battery compartment cover

removed.

●Never apply voltage to the OUTPUT terminal.

 ① Attach the output plug to a connection lead so that the output

voltage can be connected to a recording device such as a chart
recorder.