KYORITSU 2009R User Manual

Page 22

Advertising
background image

̶ 20 ̶

6-6 Continuity Check (400Ω range fixed)

 ※ The continuity check mode is enabled by pressing the mode

selector switch on resistance range. " " is indicated on the
display to show the instrument in the continuity check mode. The
buzzer beeps, if the resistance under test is 20.0Ω or less.

DANGER

●Never use the instrument on an energized circuit.
● Do not make measurement with the battery compartment cover

removed.

● Keep your fingers and hands behind the barrier during

measurement.

 ①Set the function selector switch to the " Ω" position.

 ② Slide the terminal cover to the left to disclose the V/Ω and COM

terminals. Plug the red test lead into the V/Ω terminal and the
black test lead into the COM terminal.

 ③ With the tip of the test leads shorted together, press the Zero

Adjust/Reset button to offset the resistance of the test leads.

 ④ Press the mode selector button once to enter from the normal

mode to the continuity check mode. " " should be indicated
on the display.

 ⑤ Connect the tip of the test leads to the circuit under test. If the

resistance is 20.0Ω or less, the buzzer beeps.

6-7 Frequency Measurement

 ● On ACA or ACV range, the frequency of the current or voltage

under test can be counted and shown on the display.

 ● In the frequency measurement mode, "Hz" is indicated on the

display.

 ● Trigger threshold is approx. 10V for AC voltage and approx. 40A

for AC current.