KYORITSU 2009R User Manual

Page 19

Advertising
background image

̶ 17 ̶

 ① Set the function selector switch to the " ∼ A" position. “AC"

should be shown on the upper left corner of the display.

 ② Press the trigger to open the transformer jaws and clamp them

onto a single conductor and take the reading on the display. The
most accurate reading will be obtained by keeping the conductor
at the center of the transformer jaws.

NOTE
 ● During current measurement, keep the transformer jaws fully

closed. Otherwise, accurate measurements cannot be taken.

   Maximum conductor size is 55㎜ in diameter.
 ● Zero adjustment is not necessary in AC current measurement.
 ● When the current under test measures 3% of the range or less,

or the frequency of the current is low, "LoHz" is indicated on the
display.

6-3 DC Voltage Measurement

DANGER

● Never use the instrument on a circuit above 1000VDC. This may

cause electrical shock hazard and damage to the instrument or
the circuit under test.

● Do not make measurement with the battery compartment cover

removed.

● Keep your fingers and hands behind the barrier during

measurement.

Red test lead

Black

test lead

Advertising