9

General specification, measuring range and accuracy

Voltage (VOLT) measurement of mains power supply

Measurement of Protective conductor resistance (RPE)

Measurement of Insulation resistance (RINS)

Measurement of Equivalent leakage current (IEL)

Measurement of Contact current (IL)

4. Specification

Measuring range
Resolution
Accuracy

180 ~ 260V AC
1V

±

(2%rdg+3dgt)

Rating
Measuring range
Resolution
Open-circuit voltage
Measuring current
Accuracy

20

Ω

+ and 20

Ω

- (Bi-polarity test)

0 ~ 19.99

Ω

10m

Ω

±

5.0

±

0.4V DC

Within 200 ~ 250mA DC (when measuring 0 ~ 2

Ω

)

±

(2%rdg+3dgt)

Rating
Measuring range
Resolution
Output voltage at open circuit
Short-circuit current
Rated current
Accuracy

500V / 200M

Ω

0~19.99M

Ω

/ 199.9M

Ω

(2auto ranges)

10k

Ω

/100k

Ω

500V DC +50% -0%

14mA DC or less
Within 1~1.5mA DC under the load of 0.5M

Ω

±

(2% rdg+3dgt)

Rating
Measuring range
Resolution
Open-circuit voltage
Short-circuit current
Accuracy
Internal resistance (RA Meter)

20mA AC
0 ~ 19.99mA AC
10

μ

A

30V

±

5V AC

35mA AC or less

±

(2% rdg+3dgt)

2k

Ω

Rating

Measuring range

Resolution

Accuracy

Internal resistance (RA Meter)

2mA AC

0 ~ 1.999mA AC

1

μ

A

±

(2% rdg+3dgt)

2k

Ω