KYORITSU 6202 User Manual

Page 8

Advertising
background image

6

3

Terminal block / Terminal for mains cord

The supplied mains cord can be connected to this terminal.

4

Terminal block / L/N terminal

This terminal wired in parallel to two short-circuited Phase terminals at the
test socket (1).
The DUT Phase conductors can be connected to this L/N terminal and
measured, if the DUT is not equipped with a mains plug.

5

Terminal block / PE/2mA terminal

Contact current at exposed conductive parts which are not connected with
protective conductor can be measured by connecting RPE-Test Lead. In
addition, this terminal is wired in parallel to the protective conductor
terminal at the test socket (1), so the DUT protective conductor can be
connected to this terminal, if the DUT earth terminal is not equipped with a
mains plug. Beyond this, exposed conductive parts at the DUT must be
connected to this terminal for insulation testing and the measurement of
equivalent leakage current.

6

Terminal block / RPE-Test Lead terminal

RPE-Test Lead for resistance measurement of the protective conductor can
be connected to this terminal, and the alligator clip is to be connected with
the enclosure of the DUT.

7

LCD

Measured values are displayed.

8

LED for threshold value

The red LED lights up when limit values defined in the applicable standards
have been exceeded during the measurement of protective conductor
resistance, insulation resistance, equivalent leakage current and contact
current.

9

NULL button

NULL setting shall be done before starting the measurement of protective
conductor resistance. (Cancel the resistance of the Test Lead, inside
protective fuse and so on.)

10

Range switch

Select a range with the range switch.