10

Measurement of current consumption (ILINE) (Only for M-6202)

Threshold and display (Threshold display is dependent on the accuracy on each
measuring range.)

Reference conditions

Unless otherwise specified, this specification is dependent on following
condition.

(1) Ambient temperature

: 

23

±

5

(2) Relative humidity

: 

45 ~ 75%

(3) Position

: 

Horizontal

(4) AC power source

: 

230V, 50/60Hz

(5) Altitude

: 

2000m or less

Operating temperature and humidity range

0

~ +40

Relative humidity: 85% or less (no condensation)

Storage temperature and humidity range

-20

~ +60

Relative humidity: 85% or less (no condensation)

Rate voltage and frequency

Rated voltage

: 

230V

±

10%

Rated frequency

: 

50 ~ 60Hz

±

1%

Operating method

: 

Dual slope integration

Display

: 

3 1/2 digit in LCD, character height 17mm,

Max reading 1999 with units, symbols, etc.

Maximum rated power

Approx. 9VA

Maximum throughput of M-6202

Max. 4000VA

Rating
Measuring range
Resolution
Accuracy
Throughput

16A AC
0 ~ 15.99A AC
10mA

±

(2% rdg+3dgt)

Max. 4000VA

Range

Protective conductor resistance

Insulation resistance

Equivalent leakage current

Contact current

Threshold

LED for threshold
display

Warning buzzer
(continuous sound)

RPE >0.3

Ω

RPE >1

Ω

RINS < 0.5M

Ω

RINS < 2.0M

Ω

IEL >7.0mA
IEL >15mA
IL >0.25mA
IL >0.5mA








---


---
---

---