5.2-2 Check the mains power supply voltage

There is no power switch and the instrument is immediately ready for use.
Mains power supply voltage is displayed on the LCD. Please check the value,
and when it is from 207V to 253V, the instrument can perform correct
measurements. If the displayed value is out of above range, do not make a
measurement.

WARNING

When the voltage of mains power supply is 260V or more, "HI-V" is

displayed on the LCD and buzzer sounds (discontinuous sounds). In that
case, disconnect the mains cord of the instrument from mains power supply.

5.2-3 How to use M-7121(Optional cord)

If the mains cord M-7125 is not available, or no earthing contact socket is
available, connection of the mains power supply and the instrument can be
accomplished with the help of the optional cord.

WARNING

The mains power supply to which the instrument to be connected shall be
protected against overload with fuse or circuit breaker. The rating of the
fuse and breaker shall not exceed 16A.

Be careful not to get an electrical shock when connecting cords. This
instrument can be only connected to the commercial power supply when
the voltage between Neutral (N) and Phase (L1

L3) is 230V+10%-10%,

50/60Hz.

CAUTION

The rated current capacity of this instrument, when using the optional
cord, is 10A or less.
M-7121 Phase and Neutral probe have a fuse of 600V/10A (Fast acting
type ceramic fuse) for safety purpose.