8. Fuse replacement

When the protective fuse blows during use, please replace with new one
according to below procedure.

(1) Use a flat head screwdriver and turn it

about 45

to left and remove the fuse

cap and fuse.

(2) Remove the fuse from the fuse cap

and replace it with new one.

(3) Install the fuse cap and fuse again. At

that point, the screwdriver groove
shall be at about 45

turned to left

from the initial position. Use the
flathead screwdriver and turn it to
right.

(The screwdriver groove will stop at the

horizontal position.)

WARNING

Be sure to remove mains cord from the instrument before replacing fuse.

The fuse that user can replace is this fuse only. Never attempt to perform
the other repairing.

CAUTION

Please use the specified fuse (Fast acting type ceramic fuse :
600V/500mA -

Φ

6.35

×

32mm).

For the specified fuse, purchase it by yourself or order it from our agency.