1 calibration of the output level – Anritsu AWGN MU368060A User Manual

Page 28

Advertising
background image

Section 5 Sample Calibration

5-2

5.1 Calibration of the Output Level

5.1.1 Adjusting the output level for the signal generator

Calibration of the output level at the RF connector terminal for the signal
generator is explained here.
A sample power setting with a 7.68 MHz noise signal bandwidth and –20
dBm/3.84 MHz at the RF output connector terminal is shown below.

<Calibration procedure>
(1) Set the following parameters for this unit.

Digital Mod

: On

System

: Noise1

Noise Bandwidth

: 7.68 MHz

Calculate Bandwidth : 3.84 MHz

(2) Calculate the total noise power for –20 dBm/3.84 MHz.

Read the Calculated Level on the display. With the parameter set in
Step (1), this value should be –3.45 dBm. Figure out the total noise
power using the formula below:
Total power = –20 dBm – (–3.45 dB) = –16.55 dBm

(3) Connect the power meter to the RF output connector for this unit.

(4) Set the output level for this unit so that the indication value of the

power meter becomes the total power found in Step (2). This com-
pletes calibration to the –20 dBm/3.84 MHz noise level at the RF
output connector terminal.

<Level accuracy>
Consider the level accuracy for the setting of this calibration method. It
is figured out by the following formula:
Level accuracy

= Band calculated accuracy (Calculated Level accuracy) + power me-
ter accuracy
= ±0.6 dB (when the noise band is 7.68 MHz) + power meter accu-
racy

Anritsu

Mキ3681A

Anritsu

POWER METER ML4803A