― 17 ―

③Connect the tip of the red and black test leads to the circuit under test

and take the reading on the display.

NOTE

● When the voltage under test measures 3% of the full scale or less,

or the frequency of the voltage is low, "LoHz" is indicated on the
display.

6−5 Resistance Measurement

DANGER

●Never use the instrument on an energized circuit.
●Do not make measurement with the battery compartment cover

removed.

●Keep your fingers and hands behind the barrier during measurement.

①Set the function selector switch to the "

Ω" position.

②Slide the terminal cover to the left to disclose the V/Ω and COM

terminals. Plug the red test lead into the V/Ω terminal and the black
test lead into the COM terminal.

③With the tip of the test leads shorted together, press the Zero

Adjust/Reset button to offset the resistance of the test leads.

④Connect the tip of the test leads to the circuit under test and take the

reading on the display.