― 24 ―

②Slide the terminal cover to the right to disclose the OUTPUT terminal

and insert the output plug into the terminal. Make connection to the
recording device.

③When the plug is inserted into the output terminal, auto-range function

is cleared.
Set the range depending on the state of Data Hold Switch.

Data Hold Switch OFF 400A range
Data Hold Switch ON 2000A range

Note:After measurement, be sure to return Data Hold Switch to OFF

position.

④Set the function selector switch to the desired position (ACA or DCA)

and follow appropriate measurement instructions.

NOTE

● During current measurement, keep the transformer jaws fully

closed. Otherwise, accurate measurement cannot be made. The
maximum measurable conductor size is approx. 55mm in
diameter.

● Zero adjustment is not necessary on AC current range.
● On DC current range, the Zero Adjust/Reset button may not

completely zero adjust the output voltage from the OUTPUT
terminal. In this case, make zero adjustment on the recording
device.

● Connecting the plug to the OUTPUT terminal disables the Sleep

function. The function is enabled on removing the plug from the
terminal.