KYORITSU 2009A User Manual

Page 9

Advertising
background image

― 7 ―

●Operating System

Dual Integration

●Display

Liquid crystal display with a maximum
count of 4000 plus annunciators

●Overrange Indication

"OL" is shown on the display

●Response Time

Approx. 2 seconds.

●Sample Rate

About 3 times per second.

●Location for use

Indoor use, Altitude up to 2000m

●Temperature and Humidity

23 +/-5℃, relative humidity up to 75%

for Guranteed Accuracy without condensation

●Operating Temperature

0-40℃, relative humidity up to 85%

and Humidity

without condensation

●Storage Temperature

-20-60℃, relative humidity up to 85%

and Humidity

without condensation

●Power Source

6F22(DC9V) or equivalent battery

●Current Consumption

Approx. 15mA max.

●Sleep function

Automatically powered down in about 30
minutes after the last switch operation
(power consumption: about 200μA)

●Overload Protection

DC/AC current ranges: 2400A AC for 10sec
DC/AC voltage ranges: 1200V AC/DC for 10sec
Resistance range: 600V AC/DC for 10sec

●Withstand Voltage

5500V AC, 50/60Hz for 1 minute
between electrical circuit and housing
case or metal part of the jaws

●Insulation Resistance

10MΩ or greater at 1000V between
electrical circuit and housing case or
metal part of the jaws

●Conductor Size

Approx. 55㎜ diameter max.

●Dimensions

250(L)x105(W)x49(D)mm

●Weight

About 530g

●Accessories

Test leads M-7017
6F22 battery
Carrying case M-9094
Instruction manual
Recorder output Plug M-8201

●Optional Accessories

Multi-Tran M-8008
Recorder M-5100A, etc.
Output Lead M-7014.

Advertising