User configurable settings and functions, Function overview, 4user configurable settings and functions – ROTRONIC HF65 User Manual

Page 12: 1 function overview

Advertising
background image

E-M-HF63_65-V1_11

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

Instruction Manual

Document Type

HygroFlex HF63/HF65 Humidity Temperature
Transmitters: User Guide

Document title

Page 12 of 34

© 2008; Rotronic AG E-M-HF63_65-V1_11

4

User configurable settings and functions

The HF6 ships configured as specified on the customer order. Models with analog outputs can be installed
and used just as any conventional humidity and temperature transmitter and most users will never need to use
the HF6 configurable settings and functions. Models with a digital interface generally require some
configuration by the user.

Making use of the HF6 configurable settings and functions is entirely up to the user and the appropriate
settings depend on the user application. We have provided below a short description of the HF6 functions and
also indicated the factory default settings.

4.1 Function overview

MEASUREMENT ACCURACY AND RELIABILITY

AirChip 3000 Functions

Description

► Humidity / temperature adjustment

o

1-point or multi-point humidity calibration or adjustment

o

1-point or 2-point temperature calibration or adjustment

o

Generate a time stamp for calibrations and adjustments

o

Retain and view last adjustment date and adjustment values

o

Generate calibration and adjustment protocols

► Automatic humidity sensor test
and optional drift compensation

Tests the humidity sensor for drift caused by contaminants and
can be used to automatically apply a correction. The test is
automatically carried out at regular intervals of time. Can be
configured, enabled, or disabled

The humidity sensor status can be verified either with the HW4
software or with the instrument display (if available) and is
shown as Good, SQ-tuned (corrected for drift) or Bad (defective)

► Data recording

The data recording function differs from a true data logging
function in the sense that the AirChip 3000 does not time stamp
the data. This data recording function can be used to investigate
events such as a sensor malfunction as well as to retrieve data
that would otherwise be lost

o

Start or stop data recording - up to 2000 value pairs (%RH

and temperature). Starting a recording session erases all
previously recorded data

o

The recording mode and log interval can be specified

o

When the device is powered off, the recording session is

paused but not ended As long as the recording session has
not been ended, the device automatically resumes recording
data when powered up again

o

The recorded data can be downloaded to a PC with the

HW4 software, time stamped and viewed

MEASUREMENT LOOP VALIDATION

AirChip 3000 Functions

Description

► Simulator mode

Used to make the HF6 generate fixed values for the humidity,
temperature and calculated parameter. Can be configured,
enabled or disabled