Technical data, Specifications, 10 technical data – ROTRONIC HF65 User Manual

Page 27: 1 specifications

Advertising
background image

E-M-HF63_65-V1_11

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

Instruction Manual

Document Type

HygroFlex HF63/HF65 Humidity Temperature
Transmitters: User Guide

Document title

Page 27 of 34

© 2008; Rotronic AG E-M-HF63_65-V1_11

10 Technical data

10.1 Specifications

General

HF63

HF65

Device type

Humidity temperature transmitter
with analog output signals

Humidity temperature transmitter with
digital output

Circuit type

3-wire

Mechanical configuration types

D and W


Power supply and connections

HF63

HF65

Supply voltage (VDD)

15…40 VDC or 12…28 VAC

85…265 VAC

5…40 VDC or 12…28 VAC

85…265 VAC

Nominal current consumption
(low voltage models)

< 50 mA

Model with USB interface: 50 mA

Model with Ethernet interface: 300 mA

Cable fittings

Terminal block and M16 cable grip or
½” conduit adapter

Connector (USB or Ethernet), terminal
block (power supply and RS-485) and
M16 cable grip or ½” conduit adapter

Polarity protection
(low voltage models)

Protective diode on V+


Humidity measurement

HF63

HF65

Sensor ROTRONIC

Hygromer

®

IN1

Measuring range

0…100 %RH

Measurement accuracy at 23 °C

±1.0 %RH

Repeatability 0.3

%RH

Long term stability

< 1 %RH / year

Sensor time constant

Typical 10 sec, 63% of a 35 to 80 %RH step change (1m/sec air flow at sensor)


Temperature measurement

HF63

HF65

Sensor

Pt100 RTD, IEC 751 1/3 class B

Measuring range limits

-100…150 °C / -148…302 °F

Measurement accuracy at 23 °C

±0.2 °C

Repeatability 0.05°C
Long term stability

< 0.1°C / year

Sensor time constant

Typical 4 sec, 63% of a step change (1m/sec air flow at sensor)


Calculated parameters

HF63

HF65

Psychrometric calculations

Dew or frost point (user configurable option)