UEi Test Instruments KM9106/P User Manual

Page 46

Advertising
background image

KM9106 Operators Manual

46

To access timed purge enter the following :-


Enter the code 0000 at the following display :-

ENTER SERVICE CODE

-- -- -- -- > 0 0 0 0 < -- -- ---

The following screens will be displayed, enter NO to both :-

CALIBRATE SENSOR

NO

PRINT MEMORY ?

NO


The purge sequence now begins, by using the arrow keys select YES to
‘SET PURGE ?’ and press .

SET PURGE ?

YES

Enter the purge duration and similarly the purge interval by changing each character.
Note the line below the zero indicates the number to be changed. Enter when correct.
If a time of less than the minimum or greater than the maximum then these will set as
the default values.

PURGE DURATION 05

<05 . . . . . . . . . . . . 09>

PURGE INTERVAL 30
<09 . . . . . . . . . . . . 99>

SET/CAL

ENTER

SET/CAL

ENTER

ENTER

Advertising