TECHNICAL DATA

This chapter provides the technical data of the Product.

Specifications

Property

Description / Value

Use Life/Shelf Life

Records 55 days of data, with a 9 month shelf life. Longer
recording available via custom programming.

Start Up Delay

Standard 30 minutes or configurable 0 to 1275 minutes

Report Time

Time is based on military time and reconciled with UTC

Alert Ranges

No Alerts, Above -20 °C, 2 °C to 8 °C, 15 °C to 25 °C, or
customizable between -30 °C to +60 °C

Alert Delay

Standard 30 minutes or configurable (requires special order pre-
configured product.)

Alarm Configurations Standard Cumulative or configured for Consecutive alerts
Alarm Display

Data Light emitting diodes (LED) signify temperature excursion

Temperature sampling
frequency

Five minutes

Temperature Resolution ± 0.1 °C
Temperature Accuracy ±0.5 °C (0.9 °F)
Calibration

Silicon-based temperature sensor calibrated with NIST-traceable
standards

Operating temperature
range

-30 °C to 60 °C (-22 °F to 140 °F)

Dimensions

• 45mm width x 66mm length unpouched;
• 61mm width x 85mm length pouched
• Weight 17 grams

Plastic Pouch/Housing Water resistant, IP 65 rating

Table 3

Vaisala Cold Chain Logger CCL100 Specification